GENESISXP IMPLANTABLE PULSE GENERATOR
Report
- Report Number
- 1627487-2010-02319
- Event Type
- Malfunction
- Date Received
- September 29, 2010
- Date of Event
- January 26, 2010
- Report Date
- January 27, 2010
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL: METHOD: DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: A REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT, HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HER SCS SYSTEM ON (B)(6) 2004. IT WAS REPORTED THE PT LOST STIMULATION AND COMMUNICATION FUNCTIONALITY WITH THE IPG. THE IPG WAS EXPLANTED AND REPLACED ON (B)(6) 2010. F/U ON THE PT FOUND NO FURTHER ISSUES REPORTED. THE IPG WAS NOT RETURNED TO ANS FOR EVAL. NO FURTHER INFO IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENESISXP IMPLANTABLE PULSE GENERATOR | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3644 | 24195A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |