EON MINI IMPLANTABLE PULSE GENERATOR
Report
- Report Number
- 1627487-2010-02227
- Event Type
- Malfunction
- Date Received
- September 29, 2010
- Date of Event
- January 11, 2010
- Report Date
- January 11, 2010
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVALUATION METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT, HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HER SCS SYSTEM ON (B)(6) 2008. IT WAS REPORTED, THE PT HAD NOT USED STIMULATION FOR APPROXIMATELY ONE YEAR. WHEN SHE ATTEMPTED TO TURN THE STIMULATION BACK ON HER, PT PROGRAMMER WOULD NOT COMMUNICATE WITH HER IPG. THE IPG WAS EXPLANTED AND REPLACED ON AN UNK DATE. FOLLOW UP ON THE PT FOUND NO FURTHER ISSUES REPORTED. THE EXPLANTED IPG WAS NOT RETURNED TO ANS FOR EVAL. NO FURTHER INFO IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON MINI IMPLANTABLE PULSE GENERATOR | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3788 | 173276 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |