FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY (WITH T:CONNECT MOBILE APP)
MDR report key: 18706566
·
Received February 14, 2024
Report
- Report Number
- 3013756811-2024-21799
- Event Type
- Malfunction
- Date Received
- February 14, 2024
- Date of Event
- January 25, 2024
- Report Date
- February 14, 2024
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00389152000084
- PMA / PMN Number
- K203234
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 0
IT WAS REPORTED THAT AN ALTITUDE ALARM OCCURRED WHILE THE CUSTOMER WAS NOT OUTSIDE OF LABELED OPERATING ALTITUDE RANGE. REPORTEDLY, AN OBSTRUCTION WAS BLOCKING THE SPEAKER HOLES. ADDITIONALLY, IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 300 UNITS OF INSULIN DURING THE LOAD SEQUENCE. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUES. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 130-228 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2278211 | T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY (WITH T:CONNECT MOBILE APP) | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | W0987117 | 00389152000084 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Male |