VITALITY 2
Report
- Report Number
- 2124215-2010-16990
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Date of Event
- August 25, 2010
- Report Date
- September 8, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1047-2007 TO Z-1055-20
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS (NOVEMBER 27, 2007) ADVISORY POPULATION. THIS ISSUE IS DISCUSSED IN THE Q3 2010 PRODUCT PERFORMANCE REPORT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI) BATTERY STATUS AND WAS REPLACED WITH ANOTHER ICD. UPON RECEIPT, INITIAL LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN DEVICE LABELING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | 0148| T175| E102 |