FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 1ML ST BULK CONVENIENCE PAK

MDR report key: 18705377 · Received February 14, 2024

Report

Report Number
9610847-2024-00030
Event Type
Malfunction
Date Received
February 14, 2024
Date of Event
January 24, 2024
Report Date
May 15, 2024
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
UDI-DI
30382903097013
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

OUR QUALITY ENGINEER INSPECTED THE 28 SAMPLES SUBMITTED FOR EVALUATION. THE REPORTED ISSUE OF LEAKAGE OTHER WAS NOT CONFIRMED UPON INSPECTION AND TESTING OF THE SAMPLES. ANALYSIS OF THE SAMPLE SHOWED THAT THERE WERE NO ABNORMALITIES OR DEFECTS PRESENT. THE SAMPLES WERE TESTED, AND NO LEAKAGE WAS OBSERVED. BD COULD NOT DETERMINE A MANUFACTURING RELATED ROOT CAUSE SINCE THE REPORTED DEFECT WAS NOT CONFIRMED DURING THE EVALUATION OF THE SAMPLE. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY. PRODUCTION RECORDS WERE REVIEWED, AND THESE BATCHES MEET OUR MANUFACTURING SPECIFICATION REQUIREMENTS. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. WE REGRET ANY INCONVENIENCES THIS INCIDENT MAY HAVE CAUSED YOU AND YOUR FACILITY.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYRINGE 1ML ST BULK CONVENIENCE PAK LEAKED THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY CONSUMER THAT WE ARE CURRENTLY EXPERIENCING MULTIPLE LEAKING SYRINGES WITH TWO LOT NUMBERS: (B)(6) EXP: 2027-09-30 AND (B)(6) EXP: 2027-10-31. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. MEMBER: (B)(4). VENDOR: BECTON DICKINSON (BD) CANADA INC. PRODUCT: SYRINGE, IV, STERILE, 1 ML, PHARMACY BULK TRAY PROVINCE: (B)(6) LOT NUMBER: 3212533 EXPIRY: 2027-10-31 PROBLEM DESCRIPTION: WE ARE CURRENTLY EXPERIENCING MULTIPLE LEAKING SYRINGES WITH TWO LOT NUMBERS: (B)(6) EXP: 2027-09-30 AND (B)(6) EXP: 2027-10-31.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
442741 BD SYRINGE 1ML ST BULK CONVENIENCE PAK SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 3212533 30382903097013

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown