FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1870511
·
Received October 6, 2010
Report
- Report Number
- 1824206-2010-10317
- Event Type
- Malfunction
- Date Received
- October 6, 2010
- Date of Event
- September 13, 2010
- Report Date
- September 13, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE EMERGENCY TREND VALVE WAS NOT WORKING ALTHOUGH THE TRENDELENBURG FUNCTION STILL WORKED. THE TECHNICIAN REPLACED THE EMERGENCY TREND VALVE TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE HEAD HI/LOW FUNCTION IS SLOWLY DRIFTING DOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |