FDA Adverse Event Malfunction Summary report: N

COOL-TIP SINGLE ELEC 15CM X 1 CM KIT

MDR report key: 1870453 · Received October 6, 2010

Report

Report Number
1717344-2010-00678
Event Type
Malfunction
Date Received
October 6, 2010
Date of Event
September 6, 2010
Report Date
September 6, 2010
Manufacturer
COVIDIEN LP (VALLEYLAB)
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). TO DATE, THE INCIDENT SAMPLE HAS NOT BEEN REC'D FOR EVAL. IF THE SAMPLE IS REC'D, OR IF ADD'L INFO PERTINENT TO THE INCIDENT IS OBTAINED, A F/U REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING AN ABLATION PROCEDURE, WITH THE ELECTRODE IN CORRECT PLACEMENT, THE GENERATOR WOULD NOT ACTIVATE DUE TO HIGH IMPEDANCE. THE ELECTRODE WAS REPOSITIONED AND ABLATION RESTARTED, BUT THE GENERATOR STILL REGISTERED HIGH IMPEDANCE. THE ORTHOPAEDIC SURGEON CURETTED THE OSTEOID AND REPOSITIONED THE ELECTRODE AGAIN, BUT IT STILL WOULDN'T WORK. THE DOCTOR THEN INJECTED A SMALL AMOUNT OF SALINE INTO OSTEOID AND RESTARTED THE GENERATOR. THE CYCLE STARTED. AFTER 1 MIN, THE CYCLE STOPPED DUE TO A SURFACE BURN ON THE PT'S SKIN AROUND THE ELECTRODE PLACEMENT. THE ABLATION PROCEDURE WAS STOPPED, THE ELECTRODE REMOVED, AND THE SMALL BURN WAS DRESSED WITH JELONET AND MELONIN DRESSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOL-TIP SINGLE ELEC 15CM X 1 CM KIT RF ABLATION ACCESSORY GEI COVIDIEN LP (VALLEYLAB) 177909

Patients

Seq Age Sex Outcome Treatment
1 12 YR