FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1870438 · Received October 6, 2010

Report

Report Number
2531779-2010-01477
Event Type
Malfunction
Date Received
October 6, 2010
Report Date
September 6, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

ANIMAS REC'D THE PRODUCT FOR EVAL AND NOTED THE FOLLOWING INVESTIGATION RESULTS: THE LETTERING ON THE "OK" BUTTON HAS RUBBED OFF, NO DAMAGE WAS OBSERVED TO THE KEYPAD, THE BUTTONS RESPOND TO PRESSES INTERMITTENTLY. THE "UP" BUTTON CONTACT WAS FOUND TO BE MISALIGNED AND THERE WAS CONTAMINATION UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE "DOWN" BUTTON IS STICKING AND THE "OK" SYMBOL IS WORN OFF. THE RPTR CLAIMED THAT SHE HAS TO PRESS THE BUTTON MULTIPLE TIMES FOR A RESPONSE. THE RPTR DENIED ANY EXPOSURE TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1