FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1870283
·
Received October 6, 2010
Report
- Report Number
- 2531779-2010-01484
- Event Type
- Malfunction
- Date Received
- October 6, 2010
- Date of Event
- September 6, 2010
- Report Date
- September 7, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ANIMAS AND FOUND TO HAVE ALL KEYPAD BUTTONS INTERMITTENTLY RESPONDING. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER BUTTON CONTACTS. CONTAMINATION WAS OBSERVED UNDER BUTTONS DURING INVESTIGATION.
Description of Event or Problem · 1
THE PATIENT ALLEGED THAT THE BUTTONS HAD TO BE PRESSED SEVERAL TIMES BEFORE THE PUMP WOULD RESPOND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 109 YR |