FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1870283 · Received October 6, 2010

Report

Report Number
2531779-2010-01484
Event Type
Malfunction
Date Received
October 6, 2010
Date of Event
September 6, 2010
Report Date
September 7, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ANIMAS AND FOUND TO HAVE ALL KEYPAD BUTTONS INTERMITTENTLY RESPONDING. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER BUTTON CONTACTS. CONTAMINATION WAS OBSERVED UNDER BUTTONS DURING INVESTIGATION.

Description of Event or Problem · 1

THE PATIENT ALLEGED THAT THE BUTTONS HAD TO BE PRESSED SEVERAL TIMES BEFORE THE PUMP WOULD RESPOND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 109 YR