FDA Adverse Event Malfunction Summary report: N

630G INSULIN PUMP MMT-1715K 630G BLACK MG

MDR report key: 18702690 · Received February 14, 2024

Report

Report Number
2032227-2024-125776
Event Type
Malfunction
Date Received
February 14, 2024
Date of Event
January 28, 2024
Report Date
March 22, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
P150001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PUMP PASSED THE DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST. THE PUMP WAS MONITORED FOR SEVERAL DAYS AND NO FAILED BATTERY ALERT/BATTERY FAILED ALARM NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO ALARMS/ALERTS WERE NOTED. NO POWER ERROR 25 ALARM, POWER LOSS ALARM AND REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM RECORDED IN THE FORMATTED HISTORY FILE ON THE EVEN DATE. HOWEVER, LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2024 06:23:00.000 INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2024 06:23:30.000 (B)(6) 2024 06:23:59.000 (B)(6) 2024 06:24:51.000 (B)(6) 2024 06:26:43.000 (B)(6) 2024 06:27:13.000 (B)(6) 2024 22:40:00.000 (B)(6) 2024 22:47:50.000 (B)(6) 2024 22:49:12.000 (B)(6) 2024 18:27:08.000 (B)(6) 2024 18:27:48.000 FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2024 06:23:43.000 (B)(6) 2024 06:24:17.000 (B)(6) 2024 06:25:14.000 (B)(6) 2024 06:27:05.000 (B)(6) 2024 22:49:11.000 (B)(6) 2024 18:27:09.000 UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, FAILED BATTERY ALERT/BATTERY FAILED ALARM AND LOW BATTERY ALERT WERE EXPECTED SINCE THE BATTERY IN THE PUMP IS LOW ON POWER OR DOES NOT HAVE ENOUGH POWER. THE CUSTOMER MAY HAVE USED A LOW/NO POWER/DEPLETED BATTERY. NO UNEXPECTED FAILED BATTERY ALERT/BATTERY FAILED ALARM AND LOW BATTERY ALERT NOTED DURING TESTING. THERE WERE NO OTHER UNEXPECTED PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE 28-JAN-2024 IN THE FORMATTED HISTORY FILE. THE PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, FORCE SENSOR, MOTOR AND VIBRATOR¿ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A PILLOWING KEYPAD OVERLAY, A CRACKED BATTERY TUBE THREADS, A SCRATCHED CASE AND A CRACKED CASE BEHIND THE PUMP NEAR THE BATTERY TUBE COMPARTMENT. CUSTOMER ALLEGED FOR FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS NOT CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED A BATTERY-FAILED ALARM. TROUBLESHOOTING WAS PERFORMED FOR THE REPORTED EVENT. THE CUSTOMER CALLED BACK AFTER RECEIVING A REPLACEMENT BATTERY CAP. THE CUSTOMER CONTINUED TO EXPERIENCE BATTERY-FAILED ALARMS AFTER INSERTING A NEW BATTERY WITH THE NEW BATTERY CAP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND THE PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
70108 630G INSULIN PUMP MMT-1715K 630G BLACK MG AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1715K HG491M5

Patients

Seq Age Sex Outcome Treatment
1 75 YR Female