FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1870247 · Received October 1, 2010

Report

Report Number
2531779-2010-01417
Event Type
Malfunction
Date Received
October 1, 2010
Date of Event
September 17, 2010
Report Date
September 17, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. DURING EVAL THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

DURING EVAL THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION IR 1250 NA

Patients

Seq Age Sex Outcome Treatment
1 52 YR