FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1870247
·
Received October 1, 2010
Report
- Report Number
- 2531779-2010-01417
- Event Type
- Malfunction
- Date Received
- October 1, 2010
- Date of Event
- September 17, 2010
- Report Date
- September 17, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. DURING EVAL THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
DURING EVAL THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | IR 1250 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |