FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1870223 · Received October 11, 2010

Report

Report Number
2953144-2010-02445
Event Type
Injury
Date Received
October 11, 2010
Date of Event
September 8, 2010
Report Date
September 17, 2010
Manufacturer
ABBOTT VASCULAR-REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P960043
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). 2 UNUSED REPRESENTATIVE SAMPLES WITH THE SAME PART NUMBER AS THE COMPLAINT DEVICE WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. NO MANUFACTURING OR ABNORMAL OBSERVATIONS WERE DETECTED. A ROOT CAUSE FOR THE REPORTED EXPERIENCE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PERFORATION OCCURRED DURING STENTING OF THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY WITH A NON-ABBOTT STENT, WHICH REQUIRED TREATMENT WITH A GRAFTMASTER STENT. ALTHOUGH THE PERFORATION WAS TREATED SUCCESSFULLY, THE PATIENT DEVELOPED PULSELESS ELECTRICAL ACTIVITY, AND HAD A CEREBRAL VASCULAR ACCIDENT AND EXPIRED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN PULLING THE PLUNGER BACK AND REMOVING THE NEEDLES FROM THE DEVICE, ONE OF THE SUTURES WAS BROKEN. THE DEVICE WAS REMOVED AND A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. NO ADVERSE PT EFFECTS WERE REPORTED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR-REDWOOD CITY NA 91006-6H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention