VISIAN ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2010-01013
- Event Type
- Injury
- Date Received
- October 5, 2010
- Date of Event
- June 1, 2009
- Report Date
- September 14, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
- Health Professional
- N
Narratives
(B)(4). EVAL CODES, METHOD: LENS WORK ORDER SEARCH. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND ONE SIMILAR COMPLAINT WAS FOUND. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY AND WORK ORDER SEARCH, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).
RESULTS: MEDICAL REVIEW - THE ICL IS INDICATED IN PATIENTS 21-45 YEARS OF AGE. THERE IS A MUCH GREATER RISK OF CATARACT IN PATIENTS OVER 45 YEARS OF AGE POST ICL IMPLANTATION. THE PATIENT IN THIS CASE IS OVER 45 YEARS OF AGE. ANTERIOR SUBCAPSULAR CATARACT IS A LABELED COMPLICATION IN THE IMPLANTATION OF AN ICL. CATARACT EXTRACTION MAY BE NECESSARY IF VISION IS COMPROMISED, TO PRECLUDE INCREASE IN SEVERITY OF THE CONDITION. IF THE CONDITION IS NON-PROGRESSIVE, THE SURGEON MAY OPT TO LEAVE THE ICL, ESPECIALLY WHEN THERE IS NO DECREASE IN VISUAL ACUITY. IN THE US FDA CLINICAL TRIALS, APPROXIMATELY 6% TO 7% OF EYES DEVELOPED ANTERIOR SUBCAPSULAR OPACITIES AT 7 YEARS FOLLOWING ICL IMPLANTATION, BUT ONLY 1%-2% PROGRESSED TO CLINICALLY SIGNIFICANT CATARACT DURING THE SAME PERIOD. CATARACT EXTRACTION WAS PERFORMED IN THESE CASES AND VISUAL OUTCOME WAS GOOD. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE MEDICAL REVIEW, THE MOST LIKELY CAUSE OF THE EVENT IS THE OFF-LABEL USE OF THE ICL.
THE PT REPORTED THE SURGEON IMPLANTED A 12.6 MM MICL12.6 IMPLANTABLE COLLAMER LENS IN THE PT'S RIGHT EYE (OD) ON (B)(6) 2006. THE PT REPORTED HAVING LOST VISION DUE TO THE DEVELOPMENT OF A CATARACT. THE ICL REMAINS IMPLANTED.
ADDITIONAL INFORMATION RECEIVED - THE CATARACT WAS DIAGNOSED ON (B)(6) 2009. THE CATARACT HAS NOT PROGRESSED AND THE ICL REMAINS IMPLANTED. THE PATIENT'S CURRENT VA IS 20/25 AND THE PROGNOSIS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VISIAN ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL CO. | MICL12.6 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male | Required Intervention | CARTRIDGE: MODEL AND LOT NUMBER UNK| FOAM TIP PLUNGER: MODEL AND LOT NUMBER UNK| INJECTOR: MODEL AND LOT NUMBER UNK| INJECTOR: MODEL AND LOT NUMBER UNK| FOAM TIP PLUNGER: MODEL AND LOT NUMBER UNK| CARTRIDGE: MODEL AND LOT NUMBER UNK |