FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6201 VOLUMETRIC INFUSION PUMP

MDR report key: 1870068 · Received October 15, 2010

Report

Report Number
6000001-2010-04063
Event Type
Malfunction
Date Received
October 15, 2010
Date of Event
September 16, 2010
Report Date
September 20, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF AN OCCLUSION ALARM. THE ROOT CAUSE WAS DETERMINED TO BE A WORN LATCH PIN ROLLER. NO REPAIRS WERE PERFORMED AS THE CUSTOMER DENIED THE SERVICE ESTIMATE. A SERVICE HISTORY REVIEW DETERMINED THAT THIS DEVICE HAS NOT PREVIOUSLY BEEN SERVICED BY BAXTER. A DEVICE HISTORY REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER AND THE EVALUATION IS IN PROGRESS. A FOLLOW-UP MDR WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP WHICH EXPERIENCED A FALSE OCCLUSION ALARM UPON POWER UP. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPEREINCED A FALSE OCCLUSION ALARM UPON POWER UP. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE. THE CURRENT SOFTWARE VERSION OF THIS DEVICE IS UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1