BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE
Report
- Report Number
- 9617032-2024-00183
- Event Type
- Malfunction
- Date Received
- February 13, 2024
- Date of Event
- January 15, 2024
- Report Date
- March 14, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 50382903643146
- PMA / PMN Number
- K022426
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY MATERIAL #: 364314. LOT/BATCH #: 3151647. BD HAD NOT RECEIVED SAMPLES, BUT ONE (1) PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER WAS OBSERVED. A SCREW IS OBSERVED IN THE SYRINGE BARREL. ADDITIONALLY, TEN (10) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF FOREIGN MATTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE FOREIGN MATTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE, THE DEVICE CONTAINS A FOREIGN OBJECT. NO PATIENT IMPACT REPORTED.
IT WAS REPORTED THAT WHILE USING BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE, THE DEVICE CONTAINS A FOREIGN OBJECT. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 793084 | BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 3151647 | 50382903643146 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |