HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-10823
- Event Type
- Malfunction
- Date Received
- February 13, 2024
- Date of Event
- August 30, 2022
- Report Date
- February 12, 2024
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K181216
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT. THE ROOT CAUSE OF THE VENTILATOR WAS A DEFECTIVE EXPIRATION VALVE AND CHECK VALVE ASSEMBLY. THERE WAS NO PATIENT OR USER HARM. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT.
TIGHTNESS TEST WAS WOULD NOT PASS, I COULD HEAR THE LEAK, TRIED ANOTHER EXPIRATORY VALVE AND STILL THE LEAK, REPLACED EXPIRATORY VALVE ASSEMBLY AND LEAK IS GONE, TIGHTNESS PASSES. SERVICE SOFTWARE AND FUNCTION CHECK PASS. FROM:161265:01:H15440121 TO MSP161265:01:H22160049. RELEASE VALVE DEFECTIVE WAS SHOWING ON THE VENT, IT DID GO AWAY WHEN EXPIATORY VALVE WAS REPLACED BUT THE CHECK VALVE DID FAIL DURING TESTING SO I THOUGHT IT BEST TO REPLACE IT. ONCE REPLACED NO ISSUES. SERVICE SOFTWARE AND FUNCTION CHECK ALL PASS. FROM:161192:00:1536 TO MSP161192:04:2031.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2137057 | HAMILTON MEDICAL AG | HAMILTON-T1 | CBK | HAMILTON MEDICAL AG | HAMILTON-T1 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |