PUMP 1886 780G OUS BLE PUMP MG/DL
Report
- Report Number
- 2032227-2024-124569
- Event Type
- Malfunction
- Date Received
- February 13, 2024
- Date of Event
- January 27, 2024
- Report Date
- May 16, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. NO BLANK DISPLAY NOTED. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY DOWNLOADED PUMP TRACES AND HISTORY FILE USING THUMP. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS CONFIRMED DUE TO MAIN LCD TEST FAILURE (0X00000049) IN THE BOOTLOADER. SUSPECTING HW ISSUE. PLEASE SEE BELOW FOR PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE (B)(6) 2024 IN THE FORMATTED HISTORY FILE. LOSTSENSOR1ALERT (780) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 01/21/2024 03:48:00.000; 01/23/2024 21:27:00.000; 01/25/2024 01:19:00.000; 01/25/2024 01:29:00.000; 01/25/2024 06:18:00.000; 01/25/2024 06:28:00.000; 01/26/2024 02:13:00.000; 01/26/2024 02:23:00.000; 01/26/2024 07:52:00.000; 01/26/2024 07:52:20.000. LOSTSENSOR2ALERT (781) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 01/23/2024 21:43:00.000. UNABLE TO TEST FOR LOST SENSOR ALERT DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. LOST SENSOR ALERT WAS UNKNOWN. PUMP ERROR 63 ALARM (VARIABLEINFO = 15) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 01/25/2024 06:28:19.000 AND 01/25/2024 06:38:00.000 PUMP ERROR 63 ALARM (VARIABLEINFO = 15) WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE, SUSPECTED ON HW. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 01/21/2024 03:20:25.000, 01/21/2024 03:20:26.000; 01/25/2024 06:48:48.000; 01/26/2024 22:20:34.000, 01/26/2024 22:30:00.000; 01/26/2024 22:32:41.000, 01/26/2024 22:40:16.000; 01/26/2024 23:00:04.000. POWER LOSS ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 01/26/2024 22:37:43.000. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 01/21/2024 03:20:25.000; 01/25/2024 06:28:22.000; 01/25/2024 06:48:28.000; 01/26/2024 22:31:44.000. INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, FAILED BATT ALERT/BATTERY FAILED ALARM AND POWER LOSS ALARM WERE EXPECTED SINCE THE BATTERY IN THE PUMP IS LOW ON POWER. THE CUSTOMER MAY HAVE USED A LOW POWER/DEPLETED BATTERY. UNABLE TO TEST FOR FAILED BATT ALERT/BATTERY FAILED ALARM AND POWER LOSS ALARM DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. FAILED BATT ALERT/BATTERY FAILED ALARM AND POWER LOSS ALARM WERE UNKNOWN. THE PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE PCBA 1, PCBA 2, FORCE SENSOR AND MOTOR ASSEMBLY NOTED. NO CORROSION OR MOISTURE DAMAGE FOUND ON THE VIBRATOR ASSEMBLY NOTED. THE MOLDING KNIT LINE DAMAGE (NEAR THE BELT CLIP PLATE) WAS ACCEPTABLE PER CASE ASSEMBLY COSMETIC SPECIFICATION, (B)(4). THE SC1 CAP LOCKS PROPERLY INTO THE RESERVOIR COMPARTMENT. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A SCRATCHED CASE. BLANK DISPLAY WAS NOT CONFIRMED. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS CONFIRMED DUE TO CORROSION ON THE PCBA 1 AND PCBA 2. ALSO, PUMP ERROR 63 ALARM (VARIABLEINFO = 15) WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE DUE TO CORROSION ON THE PCBA 1 AND PCBA 2. DURING VISUAL INSPECTION, CORROSION WAS ALSO FOUND ON THE FORCE SENSOR AND MOTOR ASSEMBLY NOTED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BLANK DISPLAY. TROUBLESHOOTING WAS PERFORMED AND IT WAS FOUND THAT DISPLAY DID NOT RETURN AFTER PUMP RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE DEVICE. THE PRODUCT WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 806172 | PUMP 1886 780G OUS BLE PUMP MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | HG6DYGQZZ |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Unknown |