FDA Adverse Event Injury Summary report: N

TITAN INFLATABLE PENILE PROSTHESIS FAMILY

MDR report key: 18691697 · Received February 12, 2024

Report

Report Number
2125050-2024-00218
Event Type
Injury
Date Received
February 12, 2024
Date of Event
January 19, 2023
Report Date
June 4, 2024
Manufacturer
COLOPLAST A/S
Product Code
FHW
PMA / PMN Number
P000006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE LOT NUMBER WAS REVIEWED FOR COMPLAINT TREND, NON-CONFORMING REPORT AND CAPA. DEVICES MET SPECIFICATION PRIOR TO RELEASE. NO TRENDS WERE NOTED FOR COMPLAINTS AND THERE WERE NO NON-CONFORMING REPORTS OR CAPAS THAT WERE CONFIRMED IN VEEVA TO BE ASSOCIATED.

Additional Manufacturer Narrative · 0

TITAN OTR PUMP, CYLINDERS 1 AND 2, RESERVOIR, AND DETACHED CONNECTOR/TUBING WERE RECEIVED FOR EVALUATION. ABRASION WAS NOTED ON BOTH EXHAUST TUBES AND THE INLET TUBE OF THE PUMP. NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE PUMP. NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH EITHER CYLINDER 1 OR CYLINDER 2. NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE RESERVOIR. NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE DETACHED CONNECTOR/TUBING. THE INFORMATION RECEIVED INDICATED THE DEVICE HAD A PUMP MALFUNCTION, BUT BECAUSE NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE RETURNED COMPONENTS, THE COMPLAINT COULD NOT BE CONFIRMED AS REPORTED. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. NO TRENDS WERE NOTED FOR COMPLAINTS AND THERE WERE NO NONCONFORMING REPORTS OR CAPAS THAT WERE CONFIRMED TO BE ASSOCIATED.

Description of Event or Problem · 0

ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A PUMP MALFUNCTION.

Description of Event or Problem · 0

ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A PUMP MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2306317 TITAN INFLATABLE PENILE PROSTHESIS FAMILY PENILE PROSTHESIS, INFLATABLE FHW COLOPLAST A/S 4215027_QSR9181022

Patients

Seq Age Sex Outcome Treatment
1 57 YR Unknown Required Intervention TITAN INFLATABLE PENILE PROSTHESIS ACCESSORIES| TITAN INFLATABLE PENILE PROSTHESIS ACCESSORIES| TITAN INFLATABLE PENILE PROSTHESIS FAMILY| TITAN INFLATABLE PENILE PROSTHESIS FAMILY