FDA Adverse Event Malfunction Summary report: N

ADVANTA 2 FRAME

MDR report key: 18690920 · Received February 12, 2024

Report

Report Number
3006697241-2024-00012
Event Type
Malfunction
Date Received
February 12, 2024
Date of Event
February 6, 2024
Report Date
February 12, 2024
Manufacturer
HILL-ROM MEXICO
Product Code
FNJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE BAXTER TECHNICIAN FOUND THE CPR CABLE WAS DISCONNECTED FROM THE CYLINDER. PER THE HILLROM SERVICE MANUAL THE ADVANTA¿ 2 SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. THIS WILL HELP MAKE SURE OF A LONG, OPERATIVE LIFE FOR THE ADVANTA¿ 2 BED. THE PREVENTATIVE MAINTENANCE WILL HELP TO REDUCE DOWNTIME DUE TO EXCESSIVE WEAR FAILURES. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. EXAMINE THE HANDLES, CABLES, AND CPR MECHANISM ON THE HEAD MOTOR. MAKE SURE THE SCREWS ARE INSTALLED AND FULLY TIGHTENED. RAISE THE HEAD SECTION TO THE HIGH POSITION, THEN ACTIVATE ONE OF THE CPR RELEASES. MAKE SURE THE HEAD SECTION LOWERS. ADJUST THE CPR CABLE AS NECESSARY. DO THE SAME TEST ON THE OTHER SIDE OF THE BED. RELEASE THE CPR HANDLES AND MAKE SURE THE MECHANISM LOCKS CORRECTLY BY OPERATING THE HEAD UP CONTROL FOR A FEW SECONDS. THE HEAD SECTION SHOULD RISE. REPLACE THE HEAD SECTION MOTOR AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN RECONNECTED THE CPR CABLE TO THE CYLINDER IN ORDER TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 0

BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE CPR BED FUNCTION WAS NOT LOWERING THE HEAD SECTION. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
804977 ADVANTA 2 FRAME BED, MANUAL FNJ HILL-ROM MEXICO P1190A000016

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown