FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 18686808 · Received February 12, 2024

Report

Report Number
9617032-2024-00170
Event Type
Malfunction
Date Received
February 12, 2024
Date of Event
January 24, 2024
Report Date
January 26, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903673921
PMA / PMN Number
K212724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D2A. COMMON DEVICE NAME: BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET D2B. MEDICAL DEVICE TYPE: FPA H.6. INVESTIGATION SUMMARY: MATERIAL #: 367392 LOT/BATCH #: 3123394 BD HAD NOT RECEIVED SAMPLES, BUT 3 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR CRACKED LUER ADAPTER WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF CRACKED LUER ADAPTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE CRACKED LUER ADAPTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET, THERE WAS BLOOD LEAKAGE AT THE TAIL OF THE BUTTERFLY WING NEEDLE DUE TO A CRACK. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397594 BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET SEE H.10 JKA BECTON, DICKINSON AND COMPANY (BD) 3123394 50382903673921

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown