FDA Adverse Event Malfunction Summary report: N

VITROS CHEMISTRY PRODUCTS PHYT SLIDES

MDR report key: 1868594 · Received October 13, 2010

Report

Report Number
1319809-2010-00009
Event Type
Malfunction
Date Received
October 13, 2010
Date of Event
September 14, 2010
Report Date
October 13, 2010
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
DIP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT IMPRECISE PHYT QUALITY CONTROL RESULTS WERE OBTAINED FROM THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE INVESTIGATION FOUND NO EVIDENCE TO SUGGEST THE RESULTS IN QUESTION WERE CAUSED BY AN ANALYZER MALFUNCTION. AN ALTERNATE PHYT SLIDE LOT HAS RESOLVED THE ISSUE.THE INVESTIGATION WAS UNABLE TO DETERMINE A DEFINITIVE ROOT CAUSE AND THE INVESTIGATION IS ONGOING. HOWEVER, THE MOST LIKELY CAUSE IS REAGENT RELATED.

Description of Event or Problem · 1

A CUSTOMER OBSERVED IMPRECISE VITROS PHYT QUALITY CONTROL RESULTS WHILE USING THE VITROS 5,1 FS CHEMISTRY SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF PATIENT SAMPLES WERE AFFECTED. PATIENT SAMPLES WERE TESTED DURING THE INTERVAL THAT THE IMPRECISE QUALITY CONTROL RESULTS WERE OBTAINED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT.THIS REPORT IS NUMBER ONE OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED.THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBERS 1163387 & 1170580/IVD 183544.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS CHEMISTRY PRODUCTS PHYT SLIDES IN-VITRO DIAGNOSTIC DIP ORTHO-CLINICAL DIAGNOSTICS 2649-0115-3902

Patients

Seq Age Sex Outcome Treatment
1