VITROS CHEMISTRY PRODUCTS PHYT SLIDES
Report
- Report Number
- 1319809-2010-00009
- Event Type
- Malfunction
- Date Received
- October 13, 2010
- Date of Event
- September 14, 2010
- Report Date
- October 13, 2010
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- DIP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT IMPRECISE PHYT QUALITY CONTROL RESULTS WERE OBTAINED FROM THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE INVESTIGATION FOUND NO EVIDENCE TO SUGGEST THE RESULTS IN QUESTION WERE CAUSED BY AN ANALYZER MALFUNCTION. AN ALTERNATE PHYT SLIDE LOT HAS RESOLVED THE ISSUE.THE INVESTIGATION WAS UNABLE TO DETERMINE A DEFINITIVE ROOT CAUSE AND THE INVESTIGATION IS ONGOING. HOWEVER, THE MOST LIKELY CAUSE IS REAGENT RELATED.
A CUSTOMER OBSERVED IMPRECISE VITROS PHYT QUALITY CONTROL RESULTS WHILE USING THE VITROS 5,1 FS CHEMISTRY SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF PATIENT SAMPLES WERE AFFECTED. PATIENT SAMPLES WERE TESTED DURING THE INTERVAL THAT THE IMPRECISE QUALITY CONTROL RESULTS WERE OBTAINED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT.THIS REPORT IS NUMBER ONE OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED.THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBERS 1163387 & 1170580/IVD 183544.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS CHEMISTRY PRODUCTS PHYT SLIDES | IN-VITRO DIAGNOSTIC | DIP | ORTHO-CLINICAL DIAGNOSTICS | 2649-0115-3902 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |