FDA Adverse Event Injury Summary report: N

INTELLIS

MDR report key: 18685079 · Received February 12, 2024

Report

Report Number
3004209178-2024-04435
Event Type
Injury
Date Received
February 12, 2024
Date of Event
January 24, 2024
Report Date
March 8, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LGW
UDI-DI
00763000315467
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID: 977C165, SERIAL# (B)(6), IMPLANTED: (B)(6) 2022 AND PRODUCT TYPE: LEAD. SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 977C165, SERIAL/LOT #: (B)(6), UBD: 02-MAR-2024 AND UDI#: (B)(4). MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 977C165 SERIAL# (B)(6) IMPLANTED: (B)(6) 2022 EXPLANTED: (B)(6) 2024 PRODUCT TYPE LEAD. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A PATIENT VIA MANUFACTURER REPRESENTATIVE WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THEIR WERE HIGH IMPEDANCES; ELECTRODE# 0-20150, #2-40000, #3-31610, #4-34880,#7-40000, #9-40000, #15-38070. REP MET WITH PATIENT BECAUSE SOMEONE REMOVED THE PROGRAM THEY GAVE HER IN (B)(6) 2023 AND SHE IS NOW GETTING NO RELIEF BECAUSE SHE CAN'T USE THE STIMULATOR BECAUSE SHE IS GETTING SHOCKED. HER PADDLE IS PLACED RETROGRADE IN THE CERVICAL SPINE, SO REP DECIDED TO DO AN IMPEDANCE CHECK ON HER SYSTEM BECAUSE THE LAST REP THAT PROGRAMMED HER MADE A NOTE THAT SHE HAD 4 ELECTRODES WITH HIGH IMPEDANCE. REP GAVE HER A PROGRAM USING THE AVAILABLE ELECTRODES. THE PATIENT CALLED REP TODAY TO LET THEM KNOW THAT THE PROGRAM THEY GAVE HER YESTERDAY WAS WORKING AND THEN TODAY SHE STARTED GETTING THE SHOCKING AGAIN, TO THE POINT SHE HAD TOTURN IT OFF. SHE MET UP WITH HCP AGAIN TODAY BECAUSE HE GOT THE XRAY BACK THAT HE ORDERED FOR HER YESTERDAY AND SHE TOLD HIM THAT SHE IS GETTING SHOCKED. HE IS REFERRING HER BACK TO ANOTHER HCP. ISSUE NOT RESOLVED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE (REP). IT WAS REPORTED THAT THE CAUSE FOR THE HIGH IMPEDANCE WAS NOT ABLE TO BE DETERMINED. REP REPORTED THAT THE PATIENT HAD A FULL SYSTEM REPLACEMENT (ANOTHER REP COVERED THE REPLACEMENT SURGERY, REP NOTED THAT THE EXPLANTED ITEMS WERE DISCARDED). REP REPORTED THAT THE ISSUE HAS BEEN RESOLVED; THE PATIENT IS GETTING GOOD PARESTHESIA COVERAGE WITHOUT ANY ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1063649 INTELLIS STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR LGW MEDTRONIC PUERTO RICO OPERATIONS CO. 97715 00763000315467

Patients

Seq Age Sex Outcome Treatment
1 59 YR Female Required Intervention "SEE H11...."