FDA Adverse Event Injury Summary report: N

NEEDLE 27X1-1/4 RB

MDR report key: 18684857 · Received February 12, 2024

Report

Report Number
1911916-2024-00076
Event Type
Injury
Date Received
February 12, 2024
Date of Event
January 22, 2024
Report Date
February 15, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMI
UDI-DI
30382903051367
PMA / PMN Number
K021475
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). FOLLOW UP, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 305136 AND LOT NUMBER 2021694. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Additional Manufacturer Narrative · 0

(B)(4): INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED. MATERIALS#: 305136, BATCH#: 2021694. IT WAS REPORTED BY THE CUSTOMER THAT NEEDLE SNAPPING OFF AT HUB. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. CUSTOMER REPORTS NEEDLE SNAPPING OFF AT HUB THIS HAS HAPPENED BEFORE HOWEVER ON (B)(6) 2024 STATES GIVING INJECTION TO PATIENT AND NEEDLE SNAPPED OFF IN PATIENT BUTTOCK AND SENT TO HOSPITAL. PRODUCT#: 305136. LOT#: 2021694.

Description of Event or Problem · 0

MATERIALS#: 305136. BATCH#: 2021694. IT WAS REPORTED BY THE CUSTOMER THAT NEEDLE SNAPPING OFF AT HUB. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. CUSTOMER REPORTS NEEDLE SNAPPING OFF AT HUB. THIS HAS HAPPENED BEFORE, HOWEVER ON (B)(6) 2024, STATES GIVING INJECTION TO PATIENT AND NEEDLE SNAPPED OFF IN PATIENT BUTTOCK AND SENT TO HOSPITAL. PRODUCT#: 305136. LOT#: 2021694.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
465403 NEEDLE 27X1-1/4 RB NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKINSON 2021694 30382903051367

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention