FDA Adverse Event Malfunction Summary report: N

BD MULTITEST¿

MDR report key: 18679416 · Received February 9, 2024

Report

Report Number
2916837-2024-00015
Event Type
Malfunction
Date Received
February 9, 2024
Date of Event
January 12, 2024
Report Date
June 28, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
GKZ
UDI-DI
00382906629678
PMA / PMN Number
SEE H.10
Removal / Correction Number
RES# 94830
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE (NON-SPECIFIC STAINING CREATING UNUSUAL STAINING PATTERNS FOR CD4+CD+8 POPULATIONS) WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: DATA PROVIDED BY CUSTOMER WAS REVIEWED. RETAIN TESTING RESULTS SHOWED THE PRESENCE OF A NON-SPECIFIC STAINING. POTENTIAL CAUSE IS DETERMINED TO BE MANUFACTURING RELATED. ADDITIONAL CONTROLS HAVE BEEN PUT IN PLACE IN THE MANUFACTURING STAGE SPECIFIC TO THE SUPPLIED COMPONENT. CORRECTIVE ACTION WAS INITIATED TO ADDRESS THE ROOT CAUSE AND PREVENTIVE ACTIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

G.5. PMA / 510(K)#: K090967. G.5. PMA / 510(K)#: K170974. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT ERRONEOUS RESULTS WERE OBSERVED DURING USE WITH BD MULTITEST IVD REAGENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC, CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 69761.

Description of Event or Problem · 0

IT WAS REPORTED THAT ERRONEOUS RESULTS WERE OBSERVED DURING USE WITH BD MULTITEST IVD REAGENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER FOUND THAT AS LONG AS THIS BATCH OF LOT IS STAINED, IN THE CD45 VS SSC, CD4 VS CD8 PLOTS, THERE ARE HUGE NOISE, BUT OTHER LOTS ARE NORMAL WITHOUT ANY NOISE. VERY CLEAN CD45 STAIN UNLIKE LOT 69761

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1338921 BD MULTITEST¿ COUNTER, DIFFERENTIAL CELL GKZ BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES 69761 00382906629678

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown