NAVLOCK
Report
- Report Number
- 1723170-2024-00504
- Event Type
- Malfunction
- Date Received
- February 9, 2024
- Date of Event
- January 18, 2024
- Report Date
- February 9, 2024
- Manufacturer
- MEDTRONIC NAVIGATION, INC
- Product Code
- OLO
- PMA / PMN Number
- K124004
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H3/H6: THE TRACKER, LOT NUMBER: 211121, WAS ANALYZED. ANALYSIS DETERMINED THAT ONE SIDE TAB ON THE RETURNED TRACKER WAS MISSING. OTH ERWISE, THE TRACKER WAS IN GOOD CONDITION. THE TRACKER RECEIVED KNOWN GOOD INSTRUMENTS WITHOUT ISSUE. WITH MARKERS ATTACHED AND FULLY SEATED, THE TRACKER DISPLAYED A GOOD GEOMETRY ERROR WITH NORMAL TRACKING. CODES B01, C07, AND D02 APPLY. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THE ANCHORING LOCK OF THE NAVLOCK GREEN TRACKER WAS DAMAGED. THE EVENT OCCURRED WHILE THE PATIENT WAS UNDER ANESTHESIA AND AFTER AN INCISION WAS MADE. THERE WAS NO IMPACT ON PATIENT OUTCOME. DELAY IN SURGERY WAS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 81443 | NAVLOCK | ORTHOPEDIC STEREOTAXIC INSTRUMENT | OLO | MEDTRONIC NAVIGATION, INC | 9734734 | 211121 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |