PULSE GEN MODEL 1000
Report
- Report Number
- 1644487-2024-00124
- Event Type
- Injury
- Date Received
- February 9, 2024
- Date of Event
- January 16, 2024
- Report Date
- April 1, 2024
- Manufacturer
- LIVANOVA USA, INC.
- Product Code
- LYJ
- UDI-DI
- 05425025750405
- PMA / PMN Number
- P970003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
LIVANOVA USA, INC. SUBMITS THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION, BASED ON INFORMATION THAT LIVANOVA HAS OBTAINED, BUT MAY NOT HAVE BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION, OR A CONCLUSION BY FDA OR ANYONE ELSE, THAT THE DEVICE, LIVANOVA, OR LIVANOVA'S EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS¿ OR "MALFUNCTIONS¿. THESE WORDS ARE INCORPORATED INTO THE FDA 3500A MEDWATCH FORM BY THE FDA, AND LIVANOVA OBJECTS TO THEIR USE.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SLEEP STUDY AND PER THE SLEEP SPECIALIST THE PATIENT WAS EXPERIENCING APNEA TRIGGERED BY VNS STIMULATION. THE NEUROLOGIST NOTED THAT THE WITH THEIR CURRENT DAY/NIGHT PROGRAMMING, IT DID SUGGEST THAT THE VNS CORRELATED TO THE APNEA. THE NEUROLOGIST'S PLAN WAS TO ADJUST THE PATIENT'S DAY/NIGHT SETTINGS AND WOULD MONITOR THE PATIENT MOVING FORWARD. THE PATIENT WAS SEEN IN CLINIC AND THE NIGHTTIME SETTINGS FOR THE NORMAL OUTPUT CURRENT WAS LOWERED TO 1MA. NO OTHER RELEVANT INFORMATION HAS BEEN RECEIVED TO DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51044 | PULSE GEN MODEL 1000 | GENERATOR | LYJ | LIVANOVA USA, INC. | 1000 | 815 | 05425025750405 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Female | Other |