FDA Adverse Event Malfunction Summary report: N

TMINI ARRAY FIXATION GUIDE

MDR report key: 18673298 · Received February 8, 2024

Report

Report Number
3000719653-2024-00003
Event Type
Malfunction
Date Received
February 8, 2024
Date of Event
December 19, 2023
Report Date
February 8, 2024
Manufacturer
THINK SURGICAL, INC
Product Code
OLO
PMA / PMN Number
K230202
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORT OF THE DIFFICULTY REMOVING THE ARRAY GUIDE FROM THE ARRAY FIXATION PIN COULD BE DUE TO TOLERANCE INTERFERENCE FIT WHERE THE INNER DIAMETER (ID) OF THE ARRAY GUIDE AND THE OUTER DIAMETER (OD) OF THE FIXATION PIN OR OTHER CONTRIBUTING FACTORS. THE SURGEON WAS ABLE TO SUCCESSFULLY PLACE THE FIXATION PINS AND COMPLETED THE SURGERY. NO PATIENT HAS EXPERIENCED HARM OR LONG TERM CONSEQUENCES. HOWEVER, THERE IS POSSIBILITY FOR THE FIXATION PIN TO BE UNABLE TO BE REMOVED FROM THE ARRAY GUIDE AND THE PATIENT BONE. THEREFORE, THINK SURGICAL DECIDED ON 01/31/2024 TO REPORT TO THE FDA AND VOLUNTARILY REMOVE THE ARRAY GUIDE FROM THE FIELD.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING THE PLACEMENT OF THE ARRAY FIXATION PIN USING THE ARRAY GUIDE THE PIN BOUND WITH THE ARRAY GUIDE AND THE USER HAD DIFFICULTY REMOVING THE ARRAY GUIDE FROM THE ARRAY FIXATION PIN. THE SURGEON WAS ABLE TO USE THE TRACKING ARRAY BRACKET FOR PLACEMENT OF THE FIXATION PINS AND SUCCESSFULLY COMPLETED THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68291 TMINI ARRAY FIXATION GUIDE PIN GUIDE OLO THINK SURGICAL, INC 109101 23030

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown