COBAS INTEGRA CALCIUM
Report
- Report Number
- 1823260-2010-06069
- Event Type
- Malfunction
- Date Received
- October 13, 2010
- Date of Event
- September 21, 2010
- Report Date
- October 20, 2010
- Manufacturer
- ROCHE DIAGNOSTICS CORPORATION
- Product Code
- CIC
- PMA / PMN Number
- K963292
- Removal / Correction Number
- 1823260-09/27/10-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INTERNAL INVESTIGATIONS REVEALED THE PRESENCE OF PRECIPITATE IN REAGENT 2 (R2) OF THE CALCIUM REAGENT LOT. THE ROOT CAUSE FOR THE PRECIPITATION IS STILL UNDER INVESTIGATION. AN URGENT MEDICAL DEVICE REMOVAL NOTICE WAS MAILED OCTOBER 13, 2010 TO ALL CUSTOMERS TO IMMEDIATELY DISCONTINUE USE OF THE CALCIUM REAGENT LOT. CALCIUM DEVIATIONS OF > 20% ABOVE AND BELOW THE REFERENCE RANGE ARE CONSIDERED CRITICAL. THESE DEVIATIONS COULD LEAD TO NON-DETECTION OF A PATHOLOGICAL STATUS OR MAY LEAD TO UNNECESSARY THERAPEUTIC CONSEQUENCES. THE ERRONEOUS BIAS CAUSED BY THIS ASSAY LOT MAY CAUSE PATIENT SAMPLES WITH CRITICALLY HIGH OR CRITICALLY LOW CALCIUM LEVELS TO HAVE RESULTS THAT APPEAR TO BE IN THE NORMAL RANGE. THESE PATIENTS MAY BE AT HIGH RISK FOR MISDIAGNOSIS OF PATHOLOGICAL CONDITIONS ASSOCIATED WITH ABNORMAL CALCIUM LEVELS. INACCURATELY HIGH CALCIUM VALUES MAY LEAD TO AN INCORRECT DIAGNOSIS DEPENDING ON THE BLOOD CALCIUM LEVEL. IN THE WORST CASE, THE WRONG THERAPY MAY BE INITIATED. SUSPECTED HYPERCALCEMIA WOULD MEAN CONFIRMING THE CALCIUM VALUE QUICKLY, AND TREATMENT WOULD OCCUR ACCORDING TO THE CLINICAL PICTURE. A PATIENT NORMALLY SHOWS OBVIOUS SYMPTOMS IF THE CALCIUM VALUE IS >12 MG/DL. IN THE CASE OF INACCURATELY LOW VALUES, THE CALCIUM VALUE MUST BE INTERPRETED TOGETHER WITH OTHER PARAMETERS (E.G., MAGNESIUM, PHOSPHATE, AND PTH). UNNECESSARY FURTHER EXAMINATION AND BLOOD COLLECTION MAY RESULT.
THE USER RECEIVED QUESTIONABLE CALCIUM RESULTS FOR TWO PATIENT SAMPLES FROM THE COBAS C501 ANALYZER SERIAL NUMBER (B)(4). OF THE DATA PROVIDED, THE RESULTS FOR ONE PATIENT SAMPLE WERE DISCREPANT. THE INITIAL RESULT WAS 11.7 MG/DL WITH A DATA FLAG. THE SAMPLE WAS REPEATED ON THE SAME ANALYZER WITH A RESULT OF 10.5 MG/DL AND ON THE INTEGRA 400 WITH A RESULT OF 10.5 MG/DL. THE RESULT OF 11.7 MG/DL WAS REPORTED OUTSIDE THE LABORATORY AND WAS LATER CORRECTED TO 10.5 MG/DL. THE PATIENT WAS NOT ADVERSELY AFFECTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THERE WAS A BAD REAGENT PACK AND REPLACED IT WITH A REAGENT PACK FROM A DIFFERENT SHIPMENT. THE USER RAN CALIBRATION, QUALITY CONTROL AND PRECISION TESTING WITH RESULTS WITHIN SPECIFICATION.
THE CUSTOMER REPORTED THAT THE JAW COATING CAME OFF THE DEVICE AND FELL ONTO PT TISSUE. THIS OCCURRED WITH TWO DEVICES WITHIN THE SAME SURGERY. THERE WAS NO PT INJURY. ADDITIONAL DEVICE USED IN THE PROCEDURE CAN BE FOUND ON MFR REPORT # 1717344-2010-00694.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS INTEGRA CALCIUM | CRESOLPHTHALEIN COMPLEXONE, CALCIUM | CIC | ROCHE DIAGNOSTICS CORPORATION | NA | 62601901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 019 YR |