BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC
Report
- Report Number
- 3002601200-2024-00024
- Event Type
- Malfunction
- Date Received
- February 7, 2024
- Date of Event
- January 2, 2024
- Report Date
- April 26, 2024
- Manufacturer
- BD SUZHOU (MDS)
- Product Code
- FOZ
- UDI-DI
- 00382903830831
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E1. ADDRESS INFORMATION WAS NOT PROVIDED, THEREFORE, XX WAS USED AS A PLACE HOLDER. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
1. DHR/BHR REVIEW (LOT#3171583): 1) THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN JULY 2023, AND PACKAGED AT R240 PACKAGE LINE IN JULY 2023. WORK ORDER QUANTITY WAS (B)(4) EA. 2) REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. AMONG THEM, THE PRN TORQUES MEET THE OUTGOING INSPECTION REQUIREMENT. 3) REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4) THE PRN BATCH USED IN THIS BATCH OF PRODUCTS IS 3177140, REVIEW THE RAW MATERIAL INSPECTION RECORDS, NO ABNORMALITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. THE RETAINED SAMPLE OF THIS BATCH IS TAKEN FOR LEAKAGE TEST AND PRN REMOVAL TORQUE TEST, THE LEAKAGE TEST IS QUALIFIED, THERE IS NO LEAKAGE AT THE PRN, AND THE PRN REMOVAL TORQUE IS WITHIN THE PRODUCT SPECIFICATIONS. PLEASE SEE ATTACHMENT FOR THE TEST REPORTS. 4. IN THE ASSEMBLY PROCESS OF PRN, THERE ARE TORQUE AND ASSEMBLY STROKE MONITORING TO ENSURE THAT THE LUER OF PRN CAN BE ASSEMBLED INTO THE PP CONNECTOR TO A CERTAIN DEPTH AND THE THREAD HAS A GOOD FASTENING EFFECT. IF THE PRN IS NOT IN PLACE, THE EQUIPMENT ALARMS AND REMOVES THE PRODUCT. HOWEVER, ALTHOUGH PRN IS FASTENED TO THE PRODUCT DURING ASSEMBLY, IT MAY COME LOOSE IF IT IS VIBRATED DURING TRANSPORTATION. 5. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS RETAINED SAMPLE, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. AS THE DEFECT STATUS OF THE COMPLAINED SAMPLE CANNOT BE IDENTIFIED, AND THE USAGE OF THE SAMPLE IS UNKNOWN, THE ROOT CAUSE OF THE PRN INTERFACE LEAKAGE CANNOT BE CONFIRMED.
IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: PRIOR TO PUNCTURING THE PATIENT, A FLUID LEAK FROM THE HEPARIN CAP PORT WAS DETECTED.
FLUID WAS FORCED THROUGH THE DEVICE USING A SYRINGE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1440016 | BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC | INTRAVASCULAR CATHETER | FOZ | BD SUZHOU (MDS) | 3171583 | 00382903830831 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |