FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC

MDR report key: 18664131 · Received February 7, 2024

Report

Report Number
3002601200-2024-00024
Event Type
Malfunction
Date Received
February 7, 2024
Date of Event
January 2, 2024
Report Date
April 26, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830831
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1. ADDRESS INFORMATION WAS NOT PROVIDED, THEREFORE, XX WAS USED AS A PLACE HOLDER. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW (LOT#3171583): 1) THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN JULY 2023, AND PACKAGED AT R240 PACKAGE LINE IN JULY 2023. WORK ORDER QUANTITY WAS (B)(4) EA. 2) REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. AMONG THEM, THE PRN TORQUES MEET THE OUTGOING INSPECTION REQUIREMENT. 3) REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4) THE PRN BATCH USED IN THIS BATCH OF PRODUCTS IS 3177140, REVIEW THE RAW MATERIAL INSPECTION RECORDS, NO ABNORMALITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. THE RETAINED SAMPLE OF THIS BATCH IS TAKEN FOR LEAKAGE TEST AND PRN REMOVAL TORQUE TEST, THE LEAKAGE TEST IS QUALIFIED, THERE IS NO LEAKAGE AT THE PRN, AND THE PRN REMOVAL TORQUE IS WITHIN THE PRODUCT SPECIFICATIONS. PLEASE SEE ATTACHMENT FOR THE TEST REPORTS. 4. IN THE ASSEMBLY PROCESS OF PRN, THERE ARE TORQUE AND ASSEMBLY STROKE MONITORING TO ENSURE THAT THE LUER OF PRN CAN BE ASSEMBLED INTO THE PP CONNECTOR TO A CERTAIN DEPTH AND THE THREAD HAS A GOOD FASTENING EFFECT. IF THE PRN IS NOT IN PLACE, THE EQUIPMENT ALARMS AND REMOVES THE PRODUCT. HOWEVER, ALTHOUGH PRN IS FASTENED TO THE PRODUCT DURING ASSEMBLY, IT MAY COME LOOSE IF IT IS VIBRATED DURING TRANSPORTATION. 5. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS RETAINED SAMPLE, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. AS THE DEFECT STATUS OF THE COMPLAINED SAMPLE CANNOT BE IDENTIFIED, AND THE USAGE OF THE SAMPLE IS UNKNOWN, THE ROOT CAUSE OF THE PRN INTERFACE LEAKAGE CANNOT BE CONFIRMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: PRIOR TO PUNCTURING THE PATIENT, A FLUID LEAK FROM THE HEPARIN CAP PORT WAS DETECTED.

Description of Event or Problem · 0

FLUID WAS FORCED THROUGH THE DEVICE USING A SYRINGE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1440016 BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 3171583 00382903830831

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown