FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 18664122 · Received February 7, 2024

Report

Report Number
9617032-2024-00145
Event Type
Malfunction
Date Received
February 7, 2024
Date of Event
January 17, 2024
Report Date
February 13, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903673426
PMA / PMN Number
K030573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY MATERIAL #: 367342 LOT/BATCH #: 3138775 BD HAD NOT RECEIVED SAMPLES, BUT 2 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR CRACKED LUER ADAPTER WAS OBSERVED. ADDITIONALLY, (B)(4) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF CRACKED LUER ADAPTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE CRACKED LUER ADAPTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

MEDICAL DEVICE TYPE: FPA, JKA E.1 INITIAL REPORTER FACILITY NAME: (B)(6) GENERAL HOSPITAL PMA / 510(K)#: K030573, K220212 H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET THAT DUE TO BREAKAGE OF THE HOLDER CONNECTION PART OF THE WINGED NEEDLE, NEGATIVE PRESSURE WAS NOT APPLIED, AND BLOOD LEAKED FROM THE DAMAGED AREA. NO PATIENT IMPACT OR INJURY REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET THAT DUE TO BREAKAGE OF THE HOLDER CONNECTION PART OF THE WINGED NEEDLE, NEGATIVE PRESSURE WAS NOT APPLIED, AND BLOOD LEAKED FROM THE DAMAGED AREA. NO PATIENT IMPACT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1440007 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET JKA BECTON, DICKINSON AND COMPANY (BD) 3138775 50382903673426

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown