FDA Adverse Event Malfunction Summary report: N

CPAP

MDR report key: 1866169 · Received October 12, 2010

Report

Report Number
9611451-2010-00613
Event Type
Malfunction
Date Received
October 12, 2010
Date of Event
September 8, 2010
Report Date
September 8, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZD
PMA / PMN Number
K041900
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Additional Manufacturer Narrative · 1

THE FISHER & PAYKEL HEALTHCARE HC604 CPAP AND RESPIRONICS POWER CORD WERE RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. THE CPAP AND POWER CORD WERE TESTED AS FOLLOWS: THE CPAP WAS TESTED FOR FUNCTIONALITY. A FISHER & PAYKEL HEALTHCARE POWER CORD WAS ATTACHED TO THE CPAP FOR TESTING. THE CPAP POWERED ON AND PASSED ALL FUNCTIONAL TESTS. THE CPAP WAS OPENED FOR INTERNAL INSPECTION. THERE WAS NO DAMAGE TO THE INTERNAL COMPONENTS OF THE CPAP. THE RESPIRONICS POWER CORD WAS VISUALLY INSPECTED FOR DAMAGE. THE END OF THE POWER CORD, WHERE IT ATTACHES TO THE CPAP, WAS BURNT. THE METAL WIRES AND PINS APPEARED TO HAVE BURNT COMPLETELY. NO FAULT WAS FOUND WITH THE FISHER & PAYKEL HEALTHCARE HC604 CPAP DEVICE AS IT OPERATED PROPERLY DURING TESTING AND SHOWED NO SIGNS OF INTERNAL DAMAGE. THE END OF THE RESPIRONICS POWER CORD WAS COMPLETELY BURNT. FISHER & PAYKEL HEALTHCARE SUSPECTS THAT THE FIRE WAS CAUSED BY THE CONNECTOR ON THE POWER CORD. AS FISHER & PAYKEL HEALTHCARE DOES NOT MANUFACTURE NOR SUPPLY THE COMPLAINT POWER CORD, WE ARE UNABLE TO COMMENT ON THE ROOT CAUSE OF THE FIRE. FISHER & PAYKEL HEALTHCARE HAS NOTIFIED THE DISTRIBUTOR OF THE POWER CORD, RESPIRONICS, INC., OF THIS PRODUCT COMPLAINT.

Description of Event or Problem · 1

A HOSPITAL MEDICAL SUPPLIER IN (B)(6) REPORTED THAT AN HC604 CPAP WAS PLUGGED INTO A POWER SOURCE AND THEN POWERED ON. WHEN THE UNIT WAS POWERED ON, THE PATIENT OBSERVED THAT THE POWER CORD WAS ON FIRE. THE COMPLAINANT REPORTED THAT THE PATIENT HAD BEEN USING A POWER CORD WHICH WAS SUPPLIED BY RESPIRONICS RATHER THE POWER CORD THAT WAS SUPPLIED WITH THE FISHER & PAYKEL HEALTHCARE CPAP MACHINE. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CPAP BZD BZD FISHER & PAYKEL HEALTHCARE LIMITED HC604JHU 070402

Patients

Seq Age Sex Outcome Treatment
1 RESPIRONICS POWER CORD