FDA Adverse Event Injury Summary report: N

ENROUTE TRANSCAROTID NEUROPROTECTION SYSTEM

MDR report key: 18656005 · Received February 6, 2024

Report

Report Number
3014526664-2024-00023
Event Type
Injury
Date Received
February 6, 2024
Date of Event
January 10, 2024
Report Date
February 6, 2024
Manufacturer
SILK ROAD MEDICAL
Product Code
NTE
UDI-DI
00811311020829
PMA / PMN Number
K153485
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SECONDARY MDR WAS REPORTED UNDER 3014526664-2024-00040 AS THERE WERE TWO PRODUCTS ASSOCIATED WITH THIS EVENT. THE PRODUCT ASSOCIATED WITH THIS COMPLAINT WAS NOT RETURNED TO THE MANUFACTURER FOR ANALYSIS. A REVIEW OF MANUFACTURING RECORDS FOR THIS DEVICE WAS COMPLETED AND NO ISSUES WERE IDENTIFIED THAT COULD HAVE LED TO THE ADVERSE EVENT REPORTED. AS THE REPORTED EVENT MAY RELATED TO PROCEDURAL ISSUES OR THE ENROUTE TRANSCAROTID NEUROPROTECTION SYSTEM NPS), THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION. SILK ROAD MEDICAL WILL CONTINUE TO MONITOR FOR OCCURRENCES OF SIMILAR EVENTS. ALL REASONABLY AVAILABLE INFORMATION HAS BEEN PROVIDED BY THE COMPANY AT THE TIME OF SUBMISSION OF THIS REPORT. THE FIELDS THAT ARE BLANK ARE NOT AN OMISSION AND INDICATE THAT THE INFORMATION IS EITHER NOT APPLICABLE OR CURRENTLY UNAVAILABLE. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AFTER A LEFT TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE, THE PATIENT WAS APHASIC AND HAD AN ISCHEMIC STROKE. IMAGING REVEALED LEFT SIDED SCATTERED ACUTE INFARCTS (NEW WHITE LESIONS) AND LOW DENSITY MATERIAL WITHIN THE ENROUTE STENT THAT COULD BE POTENTIALLY THROMBUS. THE PATIENT'S SYMPTOMS HAVE NOT YET RESOLVED AND NO ADDITIONAL INTERVENTION WAS PERFORMED. AS THE REPORTED EVENT MAY RELATED TO PROCEDURAL ISSUES OR THE ENROUTE TRANSCAROTID NEUROPROTECTION SYSTEM NPS), THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1180317 ENROUTE TRANSCAROTID NEUROPROTECTION SYSTEM TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE NTE SILK ROAD MEDICAL SR-200-NPS 305247 00811311020829

Patients

Seq Age Sex Outcome Treatment
1 86 YR Male Disability