FDA Adverse Event Malfunction Summary report: N

SMARTSET GHV GENTAMICIN 40G

MDR report key: 18649294 · Received February 6, 2024

Report

Report Number
1818910-2024-02422
Event Type
Malfunction
Date Received
February 6, 2024
Date of Event
January 29, 2024
Report Date
February 6, 2024
Manufacturer
DEPUY IRELAND - 3015516266
Product Code
MBB
PMA / PMN Number
OUS - SIMILA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT #: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE - 3095040, LOT NUMBER - 4200494, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED. MANUFACTURING DATE: 2023-06-21 EXPIRY DATE: 2025-05-31 QUANTITY: (B)(4). THERE WERE ZERO NONCONFORMANCES ON THIS LOT. ALL QC AND MICROBIOLOGY TESTING MET SPECIFICATION. A REVIEW OF THE DHR HAS CONFIRMED THAT THERE WERE NO PROCESS ISSUES DOCUMENTED THAT COULD CONTRIBUTE TO THE EVENT DESCRIBED. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT : A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE - 3095040, LOT NUMBER - 4200494, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN OPENED, THE OUTER BAG ALSO RIPPED OPEN THE INNER BAG, SO CONTENT WAS NOT STERILE ANYMORE TO GIVE TO THE SCRUBBED NURSE; LIKE IF THE BAGS WERE STOCK TOGETHER. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2086219 SMARTSET GHV GENTAMICIN 40G BONE CEMENT : BONE CEMENT MBB DEPUY IRELAND - 3015516266 4200494

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown