VIRTUOSO DR
Report
- Report Number
- 6000144-2010-05244
- Event Type
- Injury
- Date Received
- October 12, 2010
- Date of Event
- August 2, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- Z-0128-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
IT IS REPORTED THAT THE DEVICE BATTERY VOLTAGE IS 2.62 V TODAY, AND THAT THE DEVICE LONGEVITY IS LESS THAN EXPECTED FOR CURRENT DEVICE PROGRAMMED VALUES COMPARED WITH THE PUBLISHED SPECIFICATION. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT IS REPORTED THAT THE DEVICE BATTERY VOLTAGE IS 2.62 V TODAY, AND THAT THE DEVICE LONGEVITY IS LESS THAN EXPECTED FOR CURRENT DEVICE PROGRAMMED VALUES COMPARED WITH THE PUBLISHED SPECIFICATION. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE DEVICE HAS BEEN REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Hospitalization| O| R | (B)(4) IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD |