DEXTRUS 4136
Report
- Report Number
- 1028232-2010-02034
- Event Type
- Injury
- Date Received
- October 5, 2010
- Date of Event
- July 25, 2010
- Report Date
- September 6, 2010
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.
BOSTON SCIENTIFIC CRM RECEIVED INFO THAT TWO DAYS POST IMPLANT OF THIS DEXTRUS LEAD, CARDIAC TAMPONADE WAS OBSERVED. THE TAMPONADE WAS DRAINED AND THIS DEXTRUS RIGHT VENTRICULAR LEAD WAS REPOSITIONED AT THE SAME TIME. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFO IS RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4136 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |