FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 1864653
·
Received October 12, 2010
Report
- Report Number
- 2649622-2010-10699
- Event Type
- Injury
- Date Received
- October 12, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AN X-RAY SHOWED AN INSULATION BREAK (APPARENT CONDUCTOR FRACTURE) AND HIGH THRESHOLD. IT WAS ALSO REPORTED THAT THE LEAD HAD LOW IMPEDANCE IN UNIPOLAR MODE AND INTERMITTENT MUSCLE STIMULATION. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention | SDR303 IMPLANTABLE PULSE GENERATOR| 5592 IMPLANTABLE PACING LEAD |