FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1864653 · Received October 12, 2010

Report

Report Number
2649622-2010-10699
Event Type
Injury
Date Received
October 12, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN X-RAY SHOWED AN INSULATION BREAK (APPARENT CONDUCTOR FRACTURE) AND HIGH THRESHOLD. IT WAS ALSO REPORTED THAT THE LEAD HAD LOW IMPEDANCE IN UNIPOLAR MODE AND INTERMITTENT MUSCLE STIMULATION. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention SDR303 IMPLANTABLE PULSE GENERATOR| 5592 IMPLANTABLE PACING LEAD