FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1864544 · Received October 4, 2010

Report

Report Number
2027969-2010-01649
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
September 13, 2010
Report Date
October 4, 2010
Manufacturer
ALERE SAN DIEGO, INC
Product Code
GJS
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT THE TIME COMPLAINT WAS FILED: INRATIO: 3.0, REFERENCE: 2.5, MEAN: 2.75, CONFIDENCE LIMITS: 1.7-3.8. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT INRATIO AND REFERENCE TEST RESULT COMPARISON MEET ACCURACY CRITERIA. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO FURTHER INVESTIGATION WILL BE PURSUED. NO PRODUCT IS EXPECTED TO BE RETURNED. NO PRODUCT INFORMATION WAS PROVIDED BY THE CUSTOMER. COMPLAINT ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. THE CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2010, INRATIO: 3.0, LAB: 2.5.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC 100071 NG

Patients

Seq Age Sex Outcome Treatment
1 NI