INRATIO
Report
- Report Number
- 2027969-2010-01649
- Event Type
- Malfunction
- Date Received
- October 4, 2010
- Date of Event
- September 13, 2010
- Report Date
- October 4, 2010
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- GJS
- PMA / PMN Number
- 021923
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT THE TIME COMPLAINT WAS FILED: INRATIO: 3.0, REFERENCE: 2.5, MEAN: 2.75, CONFIDENCE LIMITS: 1.7-3.8. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT INRATIO AND REFERENCE TEST RESULT COMPARISON MEET ACCURACY CRITERIA. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO FURTHER INVESTIGATION WILL BE PURSUED. NO PRODUCT IS EXPECTED TO BE RETURNED. NO PRODUCT INFORMATION WAS PROVIDED BY THE CUSTOMER. COMPLAINT ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. THE CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2010, INRATIO: 3.0, LAB: 2.5.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC | 100071 | NG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |