ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2010-16133
- Event Type
- Malfunction
- Date Received
- October 11, 2010
- Date of Event
- August 4, 2010
- Report Date
- October 11, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR (B)(4) LABORATORY, VISUAL INSPECTION INDICATED THERE WERE SET SCREW MARK ON ALL TERMINAL CONNECTORS. THE CLEAR MEDICAL ADHESIVE WASTORN/SEPARATED FROM THE GORE COVERING AT THE PROXIMAL END OF THE SPRING ELECTRODE. THE PROXIMAL SPRING WAS STRETCHED AT THIS POINT. THERE WAS A CUT THROUGH THE INSULATION 47 CENTIMETERS FROM THE IS-1 TERMINAL PIN WHICH WAS LIKELY DAMAGE FROM THE EXPLANT PROCEDURE. RESISTANCE AND PRESSURE TEST WERE USED TO TEST ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS.
AT THIS TIME, THIS LEAD HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.
--
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS ELECTIVELY REPLACED. IT WAS NOTED THAT THE PATIENT WAS NOT CAPTURING AT 6.5 VOLTS AND THE PHYSICIAN PREFERS TO USE ACTIVE FIX LEADS; THEREFORE, THE PHYSICIAN OPTED TO REPLACE THE LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | 4076| E110| 7122| 0175 |