FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 1864207 · Received October 11, 2010

Report

Report Number
2124215-2010-16093
Event Type
Malfunction
Date Received
October 11, 2010
Date of Event
August 5, 2010
Report Date
September 8, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-0047-2008 TO Z-0053-20
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

WHEN ANALYSIS IS COMPLETE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

AT THIS TIME, NO ADDITIONAL INFORMATION IS AVAILABLE AND RECORDS INDICATE THAT THIS DEVICE REMAINS IN SERVICE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, LABORATORY TECHNICIANS UTILIZED AN ENGINEERING LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

THIS DEVICE HAS BEEN EXPLANTED AND RETURNED FOR ANALSYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE HAD TRIGGERED THE ELECTIVE REPLACEMENT INDICATOR (ERI) WITH A MONITORING BATTERY VOLTAGE OF 2.52 V AND CHARGE TIME MEASUREMENTS OF 19.2 SECONDS. NO ADVERSE PATIENT EFFECTS HAVE BEEN OBSERVED.

Description of Event or Problem · 1

ANALYSIS OF THIS DEVICE IS NOW COMPLETE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T125

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention 0185| T125| 1388T| 1488T| MISMATCH