FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 1864194 · Received October 11, 2010

Report

Report Number
2124215-2010-16142
Event Type
Injury
Date Received
October 11, 2010
Date of Event
August 6, 2010
Report Date
August 28, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD NOTED SETSCREW MARKS ON BOTH TERMINALS. THE LEAD PASSED CONDUCTOR RESISTANCE, HIGH POTENTIAL, AND INSULATION PRESSURE TESTING, CONFIRMING BOTH THE CONDUCTOR AND INSULATION WERE UNCOMPROMISED. ANALYSIS WAS UNABLE TO CONFIRM THE ALLEGATION OF DISLODGEMENT.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. WHEN TESTING IS COMPLETE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS ATRIAL DISLODGED ONE DAY POST-IMPLANT. IN A REVISION PROCEDURE, THE LEAD WAS EXPLANTED AND SUCCESSFULLY REPLACED BY A NEW LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4469

Patients

Seq Age Sex Outcome Treatment
1 48 YR Life Threatening| R 1888TC| S602| 4469| 4456