FINELINE II
MDR report key: 1864194
·
Received October 11, 2010
Report
- Report Number
- 2124215-2010-16142
- Event Type
- Injury
- Date Received
- October 11, 2010
- Date of Event
- August 6, 2010
- Report Date
- August 28, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD NOTED SETSCREW MARKS ON BOTH TERMINALS. THE LEAD PASSED CONDUCTOR RESISTANCE, HIGH POTENTIAL, AND INSULATION PRESSURE TESTING, CONFIRMING BOTH THE CONDUCTOR AND INSULATION WERE UNCOMPROMISED. ANALYSIS WAS UNABLE TO CONFIRM THE ALLEGATION OF DISLODGEMENT.
Additional Manufacturer Narrative · 1
THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. WHEN TESTING IS COMPLETE, THIS REPORT WILL BE UPDATED.
Description of Event or Problem · 1
--
Description of Event or Problem · 1
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS ATRIAL DISLODGED ONE DAY POST-IMPLANT. IN A REVISION PROCEDURE, THE LEAD WAS EXPLANTED AND SUCCESSFULLY REPLACED BY A NEW LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4469 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Life Threatening| R | 1888TC| S602| 4469| 4456 |