VITALITY 2
Report
- Report Number
- 2124215-2010-16010
- Event Type
- Malfunction
- Date Received
- October 11, 2010
- Date of Event
- August 10, 2010
- Report Date
- August 18, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1047-2007 TO Z-1055-20
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- PHYSICIAN
Narratives
UPON COMPLETION OF ANALYSIS, THIS REPORT WILL BE UPDATED.
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS (NOVEMBER 27, 2007) ADVISORY POPULATION] THIS ISSUE IS DISCUSSED IN THE Q3 2010 PRODUCT PERFORMANCE REPORT.
--
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS RETURNED FOR DISPOSAL AFTER EXPLANT. NO PRODUCT PERFORMANCE ALLEGATIONS OR ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED WITH REGARD TO THIS DEVICE. ROUTINE INITIAL RETURNED DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | 0185| T177| N119| 4591 |