VITALITY 2
Report
- Report Number
- 2124215-2010-16022
- Event Type
- Injury
- Date Received
- October 11, 2010
- Date of Event
- August 6, 2010
- Report Date
- August 23, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1047-2007 TO Z-1055-20
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THIS DEVICE WILL NOT BE RETURNED FOR ANALYSIS AS IT WAS DISCARDED BY THE HOSPITAL STAFF.
AT THIS TIME THE DEVICE HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, CRM FOR ANALYSIS. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF FURTHER INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE REOPENED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WHICH WAS INCLUDED IN THE SHORTENED REPLACEMENT WINDOW ADVISORY APRIL 2007 POPULATION PRODUCT ADVISORY WAS INITIALLY COMMUNICATED ON 4/5/2007, WAS IN STORAGE MODE UPON INTERROGATION. THE PATIENT PRESENTED TO THE CLINIC WITH A HEART RATE IN THE 20'S. ELECTIVE REPLACEMENT INDICATOR (ERI) HAD BEEN REACHED APPROXIMATELY NINE MONTHS PRIOR. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Life Threatening| R | 0185| T175 |