FDA Adverse Event Malfunction Summary report: N

SYR 10ML PUMP COMPATIBLE SALINE 10ML FIL

MDR report key: 18640736 · Received February 5, 2024

Report

Report Number
1911916-2024-00056
Event Type
Malfunction
Date Received
February 5, 2024
Date of Event
January 22, 2024
Report Date
March 20, 2024
Manufacturer
BECTON DICKINSON
Product Code
FOZ
UDI-DI
30382903065470
PMA / PMN Number
K003553
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PR (B)(4) FOLLOW UP. IT WAS REPORTED THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON. AS A SAMPLE WAS NOT RETURNED, A THOROUGH SAMPLE EVALUATION COULD NOT BE COMPLETED. TO AID IN THE INVESTIGATION, ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS A SYRINGE WITH NO PACKAGING BLISTER OR TIP CAP. THE BOTTOM PART OF THE SYRINGE HAS BLOOD AND THERE IS ALSO BLOOD PAST THE STOPPER. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS PRODUCT IS SINGLE USE AND IS NOT DESIGNED TO PULL THE PLUNGER ROD BACK. AFTER THE SALINE SOLUTION HAS BEEN EXPELLED THE SYRINGE SHOULD BE DISCARDED. IF THE PRODUCT IS NOT MEETING THE CUSTOMER¿S NEEDS, IT MAY BE BENEFICIAL TO CONTACT THE LOCAL BD SALES, OR MARKETING REPRESENTATIVE, FOR ADDITIONAL PRODUCT OPTIONS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306547, LOT 3090897. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Additional Manufacturer Narrative · 0

(B)(4) : INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

UPDATED EVENT DATE 22-JAN-2024.

Description of Event or Problem · 0

MATERIAL #: 306547 BATCH#: 3090897, 3333571, 3271933. IT WAS REPORTED BY CUSTOMER THAT THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON. VERBATIM: "I HAVE HAD ANOTHER ISSUE WITH A BD SYRINGE. HERE IS A PICTURE BELOW AND YOU CAN SEE THE LOT NUMBER. WE USE THESE SYRINGES TO IRRIGATE OUR HEMODIALYSIS CATHETER. THIS PROBLEM HAPPENED IN OUR SATELLITE UNIT IN MIRAMICHI NEW-BRUNSWICK. DURING IRRIGATION, THERE WAS A BLOOD LEAK (IF YOU CAN SEE IN THE PICTURE THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON). " ADDITIONAL INFORMATION: THOSE INCIDENT DID HAPPENED ON A DAILY BASIS, AT LEASE TWICE DAILY. WE NOW HAVE THE SAME ISSUE WITH A DIFFERENT ONE. LOT NUMBER 3333571,REF (B)(4). OCCURRED DURING FLUSH OF THE CATHETER, WHEN PULLING THE PLUNGER.

Description of Event or Problem · 0

MATERIAL #: 306547 BATCH#: 3090897 IT WAS REPORTED BY CUSTOMER THAT THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON. VERBATIM: "I HAVE HAD ANOTHER ISSUE WITH A BD SYRINGE. HERE IS A PICTURE BELOW AND YOU CAN SEE THE LOT NUMBER. WE USE THESE SYRINGES TO IRRIGATE OUR HEMODIALYSIS CATHETER. THIS PROBLEM HAPPENED IN OUR SATELLITE UNIT IN (B)(6). DURING IRRIGATION, THERE WAS A BLOOD LEAK (IF YOU CAN SEE IN THE PICTURE THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON). "

Description of Event or Problem · 0

MATERIAL #: 306547, BATCH#: 3090897, 3333571, 3271933. IT WAS REPORTED BY CUSTOMER THAT THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON. VERBATIM: "I HAVE HAD ANOTHER ISSUE WITH A BD SYRINGE. HERE IS A PICTURE BELOW AND YOU CAN SEE THE LOT NUMBER. WE USE THESE SYRINGES TO IRRIGATE OUR HEMODIALYSIS CATHETER. THIS PROBLEM HAPPENED IN OUR SATELLITE UNIT IN MIRAMICHI NEW-BRUNSWICK. DURING IRRIGATION, THERE WAS A BLOOD LEAK (IF YOU CAN SEE IN THE PICTURE THERE WAS A BLOOD SPLAT OUTSIDE THE PISTON). " ADDITIONAL INFORMATION: THOSE INCIDENT DID HAPPENED ON A DAILY BASIS, AT LEASE TWICE DAILY. WE NOW HAVE THE SAME ISSUE WITH A DIFFERENT ONE. LOT NUMBER 3333571, REF 306547. OCCURRED DURING FLUSH OF THE CATHETER, WHEN PULLING THE PLUNGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1582987 SYR 10ML PUMP COMPATIBLE SALINE 10ML FIL SALINE, VASCULAR ACCESS FLUSH FOZ BECTON DICKINSON 3090897 30382903065470
287783 SYR 10ML PUMP COMPATIBLE SALINE 10ML FIL SALINE, VASCULAR ACCESS FLUSH FOZ BECTON DICKINSON 3090897 30382903065470

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown