FDA Adverse Event Malfunction Summary report: N

PULSAR MAX

MDR report key: 1864059 · Received October 11, 2010

Report

Report Number
2124215-2010-15851
Event Type
Malfunction
Date Received
October 11, 2010
Date of Event
July 7, 2010
Report Date
August 5, 2010
Manufacturer
GUIDANT CLONMEL IRELAND
Product Code
NVZ
PMA / PMN Number
D970003
Removal / Correction Number
Z-875/897-03
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, INITIAL TESTING CONFIRMED THERE WERE NO HARDWARE RESETS AND DEVICE EXCEEDED NOMINAL PREDICTED LONGEVITY. PRE DECONTAMINATION INSPECTION CONFIRMED ALL SEAL PLUGS AND SETSCREWS WERE INTACT. THE DEVICE WAS RETURNED WITH THE HEADER OFF THE PACEMAKER CASING. VISUAL OBSERVATION OF THE DEVICE HEADER CONFIRMED NO ANOMALIES. THE DEVICE HEADER WAS THEN DISASSEMBLED AND BOTH THE INNER AND OUTER SEALING RINGS WERE INSPECTED. EVIDENCE OF SILICONE TO SILICONE BONDING WERE FOUND IN THE VENTRICULAR CHANNELS OF THE INNER SEALS. BODY FLUID CONTAMINATION WAS ALSO NOTED IN THE ATRIAL BARRELS ALL THE WAY TO THE TIP SET BLOCKS. NO FURTHER TESTING WAS PERFORMED DUE TO THE DEVICE BEING RETURNED WITH THE HEADER DETACHED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT DURING THIS PACEMAKER REPLACEMENT PROCEDURE, BOTH ATRIAL AND VENTRICULAR LEADS WERE STUCK IN THE HEADER. THE ATRIAL LEAD WAS ABLE TO BE REMOVED, HOWEVER THE VENTRICULAR LEAD COULD NOT BE REMOVED. THE PHYSICIAN USED A BONE CUTTER AND PUSHED THE VENTRICULAR LEAD OUT OF THE HEADER. THE PACEMAKER WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSAR MAX IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CLONMEL IRELAND 1270

Patients

Seq Age Sex Outcome Treatment
1 65 YR 4017| S606| 1270| 4035