PULSAR MAX
Report
- Report Number
- 2124215-2010-15851
- Event Type
- Malfunction
- Date Received
- October 11, 2010
- Date of Event
- July 7, 2010
- Report Date
- August 5, 2010
- Manufacturer
- GUIDANT CLONMEL IRELAND
- Product Code
- NVZ
- PMA / PMN Number
- D970003
- Removal / Correction Number
- Z-875/897-03
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, INITIAL TESTING CONFIRMED THERE WERE NO HARDWARE RESETS AND DEVICE EXCEEDED NOMINAL PREDICTED LONGEVITY. PRE DECONTAMINATION INSPECTION CONFIRMED ALL SEAL PLUGS AND SETSCREWS WERE INTACT. THE DEVICE WAS RETURNED WITH THE HEADER OFF THE PACEMAKER CASING. VISUAL OBSERVATION OF THE DEVICE HEADER CONFIRMED NO ANOMALIES. THE DEVICE HEADER WAS THEN DISASSEMBLED AND BOTH THE INNER AND OUTER SEALING RINGS WERE INSPECTED. EVIDENCE OF SILICONE TO SILICONE BONDING WERE FOUND IN THE VENTRICULAR CHANNELS OF THE INNER SEALS. BODY FLUID CONTAMINATION WAS ALSO NOTED IN THE ATRIAL BARRELS ALL THE WAY TO THE TIP SET BLOCKS. NO FURTHER TESTING WAS PERFORMED DUE TO THE DEVICE BEING RETURNED WITH THE HEADER DETACHED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT DURING THIS PACEMAKER REPLACEMENT PROCEDURE, BOTH ATRIAL AND VENTRICULAR LEADS WERE STUCK IN THE HEADER. THE ATRIAL LEAD WAS ABLE TO BE REMOVED, HOWEVER THE VENTRICULAR LEAD COULD NOT BE REMOVED. THE PHYSICIAN USED A BONE CUTTER AND PUSHED THE VENTRICULAR LEAD OUT OF THE HEADER. THE PACEMAKER WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PULSAR MAX | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CLONMEL IRELAND | 1270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | 4017| S606| 1270| 4035 |