FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1863962 · Received October 11, 2010

Report

Report Number
1423500-2010-04288
Event Type
Malfunction
Date Received
October 11, 2010
Date of Event
September 17, 2010
Report Date
September 17, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED OPERATIONAL AND IN GOOD CONDITION. THE DEVICE FAILED THE HOMECHOICE RITE (RETURN INSTRUMENT TEST / EVALUATION) FUNCTIONAL TEST FOR THE HEATER BAG TEMPERATURE TEST. THE DEVICE PASSED THE RITE ELECTRICAL TEST. THE PRODUCT ANALYSIS LAB EVALUATED THE DEVICE. THE TEMPMON FEATURE OF THE CRT (CYCLER REMOTE TOOLBOX) SOFTWARE WAS USED TO DIAGNOSE THE HEATER SYSTEM. NO COMMUNICATION OR HEATER ISSUES WERE OBSERVED; HEATER SYSTEM WAS FUNCTIONING PROPERLY. THE CAUSE OF THE RITE HEATER BAG TEMPERATURE TEST FAILURE WAS UNDETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, THE PRODUCT ANALYSIS LABORATORY DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE TO: HEATER BAG TEMPERATURE FAILED PERFORMANCE SPECIFICATION FOR FLUID DELIVERED OUT OF SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 90 YR