BD PANEL PHOENIX NMIC-311
Report
- Report Number
- 1119779-2024-00070
- Event Type
- Malfunction
- Date Received
- February 5, 2024
- Date of Event
- January 10, 2024
- Report Date
- June 12, 2024
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 00382904494520
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
G.5. PMA/510(K)#: THE PANEL PHOENIX NMIC/ID-311 IS AN ANTIMICROBIAL RESISTANCE PANEL THAT CONSISTS OF A COMBINATION OF THE FOLLOWING 510K NUMBERS: K032299, K061355, K061327, K031912, K023444, K063824, K033560, K063573, K041384, K060217, K052269, K190905, K032655, K062944, K060444, K173252, K033362, K062207, K063811, K163637, K151320, K063301, K031530, K060447, K032567, K060257, K023634, K123404, K020322, K132674, K173523, K063486, K022129, K023858, K071623, K031699, K060447, K024153, K060214, K132909, K042932. H.3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR HIGH MIC ERTAPENEM (ETP) AND CEFTAZIDIME (CAZ) WITH ESCHERICHIA COLI WHEN USING PHOENIX PANEL NMIC-311 (CATALOG NUMBER 449452) BATCH NUMBER 3241572. THE CUSTOMER PROVIDED PANEL RETURNS, ISOLATE RETURNS, BINARY FILES AND PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE LAB REPORTS SHOW A RESISTANT RESULT FOR ETP AND CAZ WITH E. COLI WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, ONE CUSTOMER RETURNED PANEL EACH OF THE COMPLAINT BATCH WERE INOCULATED WITH EACH CUSTOMER RETURNED ISOLATE E. COLI # 15, E. COLI # 46 AND E. COLI #84 AND PLACED IN A PHOENIX M50 FOR ETP AND CAZ MIC RESULTS. IN ADDITION, ONE CONTROL PANEL FROM THE SAME MATERIAL BUT DIFFERENT BATCH WAS INOCULATED WITH CUSTOMER RETURNED ISOLATE E. COLI # 15, E. COLI # 46 AND E. COLI #84 AND PLACED IN A PHOENIX M50 FOR ETP AND CAZ MIC RESULTS. RESULTS OF THE TEST SHOW ALL PANELS RETURNED THE EXPECTED RESULTS FOR ETP AND CAZ. THIS COMPLAINT IS NOT CONFIRMED. A REVIEW OF THE BINARY FILES WAS DETERMINED NOT TO BE REQUIRED BASED ON THE RESULTS OF THE INVESTIGATION. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. A REVIEW OF COMPLAINTS REVEALED ONE OTHER COMPLAINT ON THIS BATCH, RELATED TO THIS DEFECT AND UNCONFIRMED. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.
IT WAS REPORTED THAT WHILE TESTING WITH THE BD PANEL PHOENIX NMIC-311, CEFTAZIDIME AND ERTAPENEM WERE SENSITIVE, BUT UPON REPEAT THEY WERE RESISTANT.THERE WAS NO REPORT OF PATIENT IMPACT.
IT WAS REPORTED THAT WHILE TESTING WITH THE BD PANEL PHOENIX NMIC-311, CEFTAZIDIME AND ERTAPENEM WERE SENSITIVE, BUT UPON REPEAT THEY WERE RESISTANT. THERE WAS NO REPORT OF PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1595826 | BD PANEL PHOENIX NMIC-311 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY | LON | BECTON, DICKINSON & CO. (SPARKS) | 3241572 | 00382904494520 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |