FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1863904 · Received October 11, 2010

Report

Report Number
1423500-2010-04277
Event Type
Malfunction
Date Received
October 11, 2010
Date of Event
August 23, 2010
Report Date
September 16, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE EVALUATION DID NOT CONFIRM ANY FAILURE OR MALFUNCTION OF THE DEVICE THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS DISCOVERED DURING EVALUATION. BASED ON A REVIEW OF ALL AVAILABLE THERAPY LOG DATA, THE CAUSE OF THE IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN AND FALSE EMPTY DETECT DUE TO USE ERROR: INAPPROPRIATE BYPASS OF THE CAUTION: NEGATIVE UF ALARM. LABELING REVIEW FOUND THE LABELING TO BE ADEQUATE FOR THE USE ERROR IDENTIFIED IN THIS COMPLAINT. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE IIPV. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 7. THE PATIENT'S DRAIN VOLUME WAS 4211ML. THE LARGEST PRESCRIBED FILL VOLUME WAS 1000ML, THIS MEETS IIPV CRITERIA. PRODUCT SURVEILLANCE CONTACTED THE NURSE ON (B)(6) 2010. ACCORDING TO THE NURSE SHE WAS NOT AWARE OF THIS EVENT AS THE PATIENT DID NOT REPORT ANY SYMPTOMS OF OVERFILL. THE PATIENT HAS RESUMED THERAPY. PRODUCT SURVEILLANCE NOTIFIED THE NURSE OF THE PROBABLE CAUSE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 59 YR