FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT Ø4.3 X 10 MM

MDR report key: 18637749 · Received February 4, 2024

Report

Report Number
3011649314-2024-00078
Event Type
Injury
Date Received
February 4, 2024
Date of Event
July 29, 2022
Report Date
February 25, 2025
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT DEVICE HAS BEEN RETURNED AND THE INVESTIGATION IS CURRENTLY ONGOING. COMPLAINT HISTORY AND PRODUCT HISTORY RECORDS WERE REVIEWED, AND DOCUMENTATION INDICATED THE PRODUCT MET RELEASE CRITERIA. THE ROOT CAUSE HAS NOT BEEN IDENTIFIED. ONCE THE RETURNED DEVICE HAS BEEN EVALUATED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). SECTION E1 POST OFFICE OR ZIP CODE: (B)(6). SECTION E1 TELEPHONE NUMBER: (B)(6).

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS THE DHR WAS REVIEWED FOR LOT# 6112671 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS THE STOCK PRODUCT FOR HAHN TAPERED IMPLANT LOT# 6112671 IS NOT APPLICABLE FOR REVIEW SINCE THE ISSUE IS CUSTOMER RELATED. INVESTIGATION METHODS/RESULTS THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø4.3 X 10 MM (70-1154-IMP0010) USING RADIOGRAPHIC TEMPLATE PK- (B)(6). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. BONE DEBRIS WAS OBSERVED IN THE THREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE "LOSS OF OSSEOINTEGRATION" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LOSS OF OSSEOINTEGRATION IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 012566 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU 012566 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU 012566 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU 012566 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." PER THE REPORTED INFORMATION, THERE WAS A FIT ISSUE. IFU 012566 REV 3 (HAHN TAPERED IMPLANT SYSTEM MULTI-LANGUAGE IFU) CONTAINS THE FOLLOWING STATEMENT IN THE SITE PREPARATION SECTION: "IF PLACING A HAHN TAPERED IMPLANT THAT IS 3.5 MM IN DIAMETER OR GREATER, SHAPING DRILLS ARE USED SEQUENTIALLY TO WIDEN THE OSTEOTOMY TO THE MATCHING DIAMETER. TO AVOID OVER-PREPARATION, WIDENING DRILL DIAMETERS SHOULD BE USED ONLY AS NEEDED, AND IN PROPER SUCCESSION. EACH SHAPING DRILL IS LENGTH-SPECIFIC, TO MATCH THE LENGTH OF THE PRESCRIBED IMPLANT. OSTEOTOMY DEPTH MAY BE INCREASED SEQUENTIALLY, BEGINNING WITH SHORTER DRILL LENGTHS, PROVIDED SUFFICIENT DEPTH IS ACHIEVED WITH THE FINAL DRILL. SELECT THE DESIRED SHAPING DRILL, ACCOUNTING FOR BONE DENSITY AND THE SIZE OF THE IMPLANT TO BE PLACED. WITH COPIOUS IRRIGATION, DRILL TO DEPTH. THE FINAL DRILL SHOULD CORRESPOND WITH THE MATCHING IMPLANT SIZE, AS CHARTED BELOW, WITH THE GOAL OF ACHIEVING HIGH PRIMARY STABILITY UPON IMPLANT PLACEMENT." CORRECTION - H6 MEDICAL DEVICE PROBLEM CODE.

Additional Manufacturer Narrative · 0

CORRECTED B1,B2,H1, G1 CONTINUED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

CORRECTED D4. CAPA CA-00016. MANUFACTURED REFERENCE IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT THE HAHN TAPERED LOSS OSSEOINTEGRATION DURING IMPLANT PLACEMENT ON TOOTH #37. PER THE REPORTED INFORMATION THERE ARE NO PREEXISTING MEDICAL CONDITIONS. THE PATIENT'S BONE QUALITY IS TYPE III WITH UNKNOWN ORAL HYGIENE STATUS. THE DEVICE WAS REMOVED AND REPLACED WITH ANOTHER IMPLANT; MODEL UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
278644 HAHN TAPERED IMPLANT Ø4.3 X 10 MM HAHN TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0010 6112671

Patients

Seq Age Sex Outcome Treatment
1 34 YR Female Required Intervention