ARCHITECT TOTAL B-HCG REAGENT KIT
Report
- Report Number
- 3005094123-2024-00045
- Event Type
- Malfunction
- Date Received
- February 4, 2024
- Date of Event
- December 25, 2023
- Report Date
- March 7, 2024
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- DHA
- PMA / PMN Number
- K983424
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SECTION A1 - PATIENT IDENTIFIER: SIDS = (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT TOTAL B-HCG REAGENT, LIST 07K78-35, ABBOTT IRELAND DIAGNOSTICS DIVISION, LONGFORD, IRELAND TO THE ARCHITECT I2000SR PROCESSING MODULE, LIST 03M74-02, ABBOTT LABORATORIES, IRVING, TEXAS, USA. MDR NUMBER 3016438761-2024-00142 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
THE CUSTOMER OBSERVED FALSE POSITIVE ARCHITECT TOTAL ¿-HCG RESULTS GENERATED ON THE ARCHITECT I2000SR PROCESSING MODULE FOR TWO PATIENTS. THE SAMPLES WERE NEGATIVE UPON REPEAT. THE CUSTOMER ALSO NOTED THAT THE LEVEL 1 CONTROL SOMETIMES RUN HIGH BUT CAME IN RANGE WHEN REPEATED. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2023 SID (B)(6) INITIAL RESULT = 5088.75 MIU/ML (POSITIVE). (B)(6) 2023 REPEAT RESULTS = <1.20MIU/ML, <1.20MIU/ML, <1.20MIU/ML (ALL NEGATIVE). (B)(6) 2024 SID (B)(6) INITIAL RESULT = 124.20 MIU/ML (POSITIVE). (B)(6) 2024 REPEAT RESULT = <1.20MIU/ML (NEGATIVE). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1373532 | ARCHITECT TOTAL B-HCG REAGENT KIT | SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN | DHA | ABBOTT IRELAND DIAGNOSTICS DIVISION | 55315UD00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC I2000SR INST, 03M74-02, (B)(6)| ARC I2000SR INST, 03M74-02, (B)(6) |