FDA Adverse Event Malfunction Summary report: N

THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER

MDR report key: 18636410 · Received February 2, 2024

Report

Report Number
2029046-2024-00392
Event Type
Malfunction
Date Received
February 2, 2024
Date of Event
November 16, 2023
Report Date
February 2, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
LPB
UDI-DI
10846835010183
PMA / PMN Number
P030031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. VISUAL INSPECTION AND SCREENING TEST OF THE RETURNED DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED REDDISH MATERIAL AND A HOLE IN THE SURFACE OF THE PEBAX. THE MAGNETIC AND FORCE FEATURE WERE TESTED, AND NO ERRORS WERE OBSERVED. THE FORCE VALUES AND THE VECTOR WERE OBSERVED WITHIN SPECIFICATIONS. NO FORCE ISSUES WERE OBSERVED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 31147797L NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. THE REDDISH MATERIAL INSIDE THE PEBAX COULD BE RELATED TO THE FORCE ISSUE REPORTED BY THE CUSTOMER; THEREFORE, THE CUSTOMER COMPLAINT WAS CONFIRMED. THE ROOT CAUSE OF THE PEBAX DAMAGE COULD BE RELATED TO THE MANIPULATION OF THE DEVICE DURING THE PROCEDURE. HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED. IT SHOULD BE NOTED THAT PRODUCT FAILURE IS MULTIFACTORIAL. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH BWI¿S QUALITY SYSTEM. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER FOR WHICH BIOSENSE WEBSTER¿S PRODUCT ANALYSIS LAB (PAL) IDENTIFIED A HOLE IN THE SURFACE OF THE PEBAX. INITIALLY, IT WAS REPORTED THAT THE FORCE VALUES WOULD JUMP ON THE CARTO 3 SYSTEM WHEN ABLATING. THE CABLE WAS REPLACED WITH NO RESOLUTION. THE CATHETER WAS RESEATED WITH NO RESOLUTION. WHEN THE CATHETER WAS REPLACED, THE ISSUE WAS RESOLVED. NO ADVERSE PATIENT CONSEQUENCE WAS REPORTED. THE FORCE ISSUE WAS ASSESSED AS NON MDR REPORTABLE. THE POTENTIAL RISK THAT IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH TO THE OPERATOR OR PATIENT WAS REMOTE. THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION AND PER THE EVALUATION COMPLETION ON 11-JAN-2024 THERE WAS A HOLE IN THE SURFACE OF THE PEBAX. THE HOLE IN THE PEBAX WAS ASSESSED AS MDR REPORTABLE. THE AWARENESS DATE FOR THIS REPORTABLE LAB FINDING WAS 11-JAN-2024.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
278361 THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER CARDIAC ABLATION PERCUTANEOUS CATHETER LPB BIOSENSE WEBSTER INC 31147797L 10846835010183

Patients

Seq Age Sex Outcome Treatment
1 Unknown UNK BRAND CABLE.| UNK BRAND CABLE.| UNK_CARTO 3.